Overview

Product name: Human FOXO3(Forkhead Box Protein O3) ELISA Kit
Reactivity: Human
Alternative Names: FOX-O3; FOX2; FOXO3A; FKHRL1; FKHRL1P2; AF6q21 protein; Forkhead in rhabdomyosarcoma-like 1
Assay Type: Sandwich
Sensitivity: 6.74 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample type: serum, plasma, other biological fluids
Assay length: 3.5h
Research Area: Signal transduction;Tumor immunity;Immune molecule;Developmental science;Autoimmunity;
Uniprot ID: O43524
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human FOXO3. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human FOXO3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human FOXO3, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human FOXO3 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
1000.00 2.153 2.072
500.00 1.593 1.512
250.00 1.186 1.105
125.00 0.871 0.790
62.50 0.468 0.387
31.25 0.325 0.244
15.63 0.217 0.136
0.00 0.081 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant FOXO3 and the recovery rates were calculated by comparing the measured value to the expected amount of FOXO3 in samples.
Matrix Recovery range Average
serum(n=5) 80-97% 88%
EDTA plasma(n=5) 78-92% 85%
Heparin plasma(n=5) 90-105% 97%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FOXO3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 98-103% 86-92% 87-93% 93-102%
EDTA plasma(n=5) 85-92% 79-96% 89-101% 85-97%
Heparin plasma(n=5) 86-94% 93-102% 89-97% 85-94%