Overview

Product name: Human GJb1(Gap Junction Protein Beta 1) ELISA Kit
Reactivity: Human
Alternative Names: CMTX1; CX32; GJ-B1; Connexin 32; Charcot-Marie-Tooth Neuropathy,X-Linked; GAP junction 28 kDa liver protein
Assay Type: Sandwich
Sensitivity: 0.113 ng/mL
Standard: 20 ng/mL
Detection Range: 0.32-20 ng/mL
Sample type: Tissue homogenates and other biological fluids.
Assay length: 3.5h
Research Area: Tumor immunity;Infection immunity;Neuro science;Bone metabolism;
Uniprot ID: P08034
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human GJb1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human GJb1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human GJb1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human GJb1 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
20.00 1.941 1.845
10.00 1.515 1.419
5.00 1.142 1.046
2.50 0.827 0.731
1.25 0.585 0.489
0.63 0.308 0.212
0.32 0.206 0.110
0.00 0.096 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant GJb1 and the recovery rates were calculated by comparing the measured value to the expected amount of GJb1 in samples.
Matrix Recovery range Average
serum(n=5) 82-98% 90%
EDTA plasma(n=5) 80-97% 88%
Heparin plasma(n=5) 82-94% 88%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of GJb1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-101% 82-98% 89-97% 85-93%
EDTA plasma(n=5) 85-92% 79-96% 95-103% 87-98%
Heparin plasma(n=5) 87-96% 85-97% 85-94% 87-96%