99499威尼斯(中国·WWW认证)游戏内容-Macau App Station

产品中心

人柠檬酸合酶(CS)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK4619
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human CS(Citrate Synthase) ELISA Kit
Reactivity: Human
Alternative Names: Citrate (Si)-synthase; mitochondrial
Assay Type: Sandwich
Sensitivity: 2.49 ng/mL
Standard: 400 ng/mL
Detection Range: 6.25-400 ng/mL
Sample type: Serum, plasma, tissue homogenates and other biological fluids.
Assay length: 3.5h
Research Area: Enzyme & Kinase;Metabolic pathway;Developmental science;
Uniprot ID: O75390
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human CS. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human CS. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human CS, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human CS in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
400.00 2.152 2.059
200.00 1.507 1.414
100.00 1.102 1.009
50.00 0.752 0.659
25.00 0.597 0.504
12.50 0.357 0.264
6.25 0.216 0.123
0.00 0.093 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant CS and the recovery rates were calculated by comparing the measured value to the expected amount of CS in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 82-95% 88%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CS and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 85-97% 90-99% 82-101% 93-105%
EDTA plasma(n=5) 95-104% 82-96% 92-103% 90-99%
Heparin plasma(n=5) 85-92% 83-96% 93-102% 87-98%