人脑特异性血管生长抑制因子1(BAI1)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK4950
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human BAI1(Brain Specific Angiogenesis Inhibitor 1) ELISA Kit
Reactivity: Human
Alternative Names: BAI 1; BAI1; GDAIF
Assay Type: Sandwich
Sensitivity: 12.8 pg/mL
Standard: 2000 pg/mL
Detection Range: 31.25-2000 pg/mL
Sample type: Serum, plasma, tissue homogenates and other biological fluids
Assay length: 3.5h
Research Area: Tumor immunity;Neuro science;
Uniprot ID: O94812
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human BAI1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human BAI1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human BAI1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human BAI1 in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
2000.00 2.096 2.015
1000.00 1.577 1.496
500.00 1.035 0.954
250.00 0.815 0.734
125.00 0.567 0.486
62.50 0.323 0.242
31.25 0.197 0.116
0.00 0.081 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant BAI1 and the recovery rates were calculated by comparing the measured value to the expected amount of BAI1 in samples.
Matrix Recovery range Average
serum(n=5) 80-95% 87%
EDTA plasma(n=5) 86-99% 92%
Heparin plasma(n=5) 82-95% 88%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BAI1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 98-106% 86-99% 78-95% 89-97%
EDTA plasma(n=5) 85-94% 87-96% 92-101% 82-98%
Heparin plasma(n=5) 79-92% 89-102% 87-98% 80-92%