48T | ¥2240.00 |
96T | ¥3200.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.030 | 1.954 |
5.00 | 1.532 | 1.456 |
2.50 | 1.235 | 1.159 |
1.25 | 0.751 | 0.675 |
0.63 | 0.425 | 0.349 |
0.31 | 0.375 | 0.299 |
0.16 | 0.190 | 0.114 |
0.00 | 0.076 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 91-99% | 95% |
EDTA plasma(n=5) | 89-127% | 108% |
Heparin plasma(n=5) | 84-119% | 102% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 86-101% | 93-116% | 84-90% | 95-102% |
EDTA plasma(n=5) | 88-96% | 90-105% | 81-100% | 92-112% |
Heparin plasma(n=5) | 91-107% | 85-119% | 92-109% | 92-111% |