Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
10000.00 | 0.219 | |
5000.00 | 0.433 | |
2500.00 | 0.711 | |
1250.00 | 0.959 | |
625.00 | 1.275 | |
312.50 | 1.565 | |
156.25 | 1.835 | |
0.00 | 2.316 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-92% | 85% |
EDTA plasma(n=5) | 85-99% | 92% |
Heparin plasma(n=5) | 87-99% | 91% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 92-107% | 83-99% | 80-95% | 87-96% |
EDTA plasma(n=5) | 85-92% | 79-96% | 89-100% | 85-96% |
Heparin plasma(n=5) | 91-99% | 85-97% | 85-94% | 87-96% |