Overview

Product name: CNCbl(Cyanocobalamin) ELISA Kit
Reactivity: General
Alternative Names: VB12; Vitamin B12; Cobalamin; α-5,6-Dimethylbenzimidazolyl Cobamidcyanide
Assay Type: Competitive Inhibition
Sensitivity: 49.1 pg/mL
Standard: 10000 pg/mL
Detection Range: 156.25-10000 pg/mL
Sample type: serum, plasma, tissue homogenates and other biological fluids
Assay length: 2h
Research Area: Metabolic pathway;Nutrition metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with CNCbl. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to CNCbl. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of CNCbl in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
10000.00 0.179
5000.00 0.336
2500.00 0.551
1250.00 0.897
625.00 1.236
312.50 1.457
156.25 1.863
0.00 2.219

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant CNCbl and the recovery rates were calculated by comparing the measured value to the expected amount of CNCbl in samples.
Matrix Recovery range Average
serum(n=5) 80-96% 88%
EDTA plasma(n=5) 80-92% 86%
Heparin plasma(n=5) 78-95% 87%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of CNCbl and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 83-96% 95-107% 82-96%
EDTA plasma(n=5) 83-97% 86-97% 91-99% 95-103%
Heparin plasma(n=5) 86-98% 87-102% 85-96% 87-101%