99499威尼斯(中国·WWW认证)游戏内容-Macau App Station

产品中心

高半胱氨酸(HCy)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK8121
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview 文献(6)

Product name: HCy(Homocysteine) ELISA Kit
Reactivity: General
Alternative Names: Homocysteine
Assay Type: Competitive Inhibition
Sensitivity: 40.95 ng/mL
Standard: 8000 ng/mL
Detection Range: 125-8000 ng/mL
Sample type: serum, plasma and other biological fluids
Assay length: 2h
Research Area: atherosclerosis
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with HCy. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to HCy. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of HCy in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (ng/mL) OD Corrected OD
8000.00 0.235
4000.00 0.433
2000.00 0.786
1000.00 0.959
500.00 1.275
250.00 1.579
125.00 1.835
0.00 2.375

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant HCy and the recovery rates were calculated by comparing the measured value to the expected amount of HCy in samples.
Matrix Recovery range Average
serum(n=5) 82-95% 88%
EDTA plasma(n=5) 85-97% 91%
Heparin plasma(n=5) 95-107% 91%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HCy and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-97% 83-96% 95-107% 82-96%
EDTA plasma(n=5) 81-94% 93-107% 93-102% 91-105%
Heparin plasma(n=5) 92-106% 83-94% 85-98% 85-97%