99499威尼斯(中国·WWW认证)游戏内容-Macau App Station

产品中心

人突触核蛋白β(SNCb)微量上样酶联免疫吸附检测试剂盒
一键复制产品信息
ELK8964MS
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human SNCb(Synuclein beta) Microsample ELISA Kit
Reactivity: Human
Alternative Names: Beta synuclein; SNCB; synuclein beta
Assay Type: Sandwich
Sensitivity: 6.9 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma, cerebrospinal fluid and other biological fluids
Assay Length: 3.5h
Research Area: Signal transduction;Neuro science;
Uniprot ID: Q16143
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human SNCb. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human SNCb. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human SNCb, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human SNCb in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
1000.00 2.300 2.207
500.00 1.670 1.577
250.00 1.094 1.001
125.00 0.827 0.734
62.50 0.461 0.368
31.25 0.368 0.275
15.63 0.213 0.120
0.00 0.093 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant SNCb and the recovery rates were calculated by comparing the measured value to the expected amount of SNCb in samples.
Matrix Recovery range Average
serum(n=5) 85-115% 100%
EDTA plasma(n=5) 82-93% 88%
Heparin plasma(n=5) 90-98% 94%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of SNCb and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-100% 82-120% 91-98% 83-110%
EDTA plasma(n=5) 82-101% 86-91% 89-112% 87-102%
Heparin plasma(n=5) 81-107% 84-117% 89-119% 90-111%

产品成分表

关闭

在线咨询

Online consultation

  • 在线咨询
  • 技术支持

关注微信公众号

微信扫一扫立即咨询

微信扫一扫立即咨询

关闭
Baidu
sogou