99499威尼斯(中国·WWW认证)游戏内容-Macau App Station

产品中心

乙酰辅酶A(AcCoA)酶联免疫吸附检测试剂盒
一键复制产品信息
ELK9575
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: AcCoA(Acetyl CoA) ELISA Kit
Reactivity: General
Alternative Names: AcCoA
Assay Type: Sandwich
Sensitivity: 0.116 nmol/mL
Standard: 20 nmol/mL
Detection Range: 0.32-20 nmol/mL
Sample type: Cell Lysate, Tissue Lysate
Assay length: 3.5h
Research Area: Metabolic pathway; Signal Transduction; Cancer
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to AcCoA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to AcCoA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain AcCoA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of AcCoA in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (nmol/mL) OD Corrected OD
20.00 2.096 2.028
10.00 1.429 1.361
5.00 1.185 1.117
2.50 0.848 0.780
1.25 0.453 0.385
0.63 0.323 0.255
0.32 0.213 0.145
0.00 0.068 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant AcCoA and the recovery rates were calculated by comparing the measured value to the expected amount of AcCoA in samples.
Matrix Recovery range Average
serum(n=5) 86-99% 92%
EDTA plasma(n=5) 87-99% 93%
Heparin plasma(n=5) 80-95% 87%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of AcCoA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 81-92% 87-101% 89-99% 85-95%
EDTA plasma(n=5) 93-102% 93-106% 87-102% 84-98%
Heparin plasma(n=5) 89-102% 82-96% 88-97% 85-92%
关闭

在线咨询

Online consultation

  • 在线咨询
  • 技术支持

关注微信公众号

微信扫一扫立即咨询

微信扫一扫立即咨询

关闭
Baidu
sogou