人酰基化饥饿素(A-GHRL)微量上样酶联免疫吸附检测试剂盒
一键复制产品信息
ELK9594MS
规格: 价格:
48T ¥1960.00
96T ¥2800.00

Overview

Product name: Human A-GHRL(AcylatedGhrelin) Microsample ELISA Kit
Reactivity: Human
Alternative Names: A-GHRL
Assay Type: Sandwich
Sensitivity: 9.24 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cardiovascular;Signal Transduction;Cardiovascular;Metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human A-GHRL. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human A-GHRL. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human A-GHRL, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human A-GHRL in the samples is then determined by comparing the OD of the samples to the standard curve.

标准曲线

Concentration (pg/mL) OD Corrected OD
1000.00 1.993 1.909
500.00 1.531 1.447
250.00 1.248 1.164
125.00 0.953 0.869
62.50 0.506 0.422
31.25 0.350 0.266
15.63 0.189 0.105
0.00 0.084 0.000

精密度

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

回收率

Matrices listed below were spiked with certain level of recombinant A-GHRL and the recovery rates were calculated by comparing the measured value to the expected amount of A-GHRL in samples.
Matrix Recovery range Average
serum(n=5) 91-104% 98%
EDTA plasma(n=5) 85-96% 91%
Heparin plasma(n=5) 93-102% 98%

线性

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of A-GHRL and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-99% 82-109% 83-89% 89-108%
EDTA plasma(n=5) 82-93% 81-105% 83-117% 90-113%
Heparin plasma(n=5) 80-101% 89-97% 85-106% 86-99%